Pressure Out of Range In Vacuum See
Op Manual Appendix A
Problem/Condition
Pressure did not stabilize within the range of 475 mTorr to 525 mTorr
after 5 minutes of pre-plasma, or was out of range more than 60 seconds
after the pressure stabilized.
Action
F
Check the load for wet or plastic like material that may be
out-gassing.
F
Repackage the load and restart the sterilizer.
If the problem persists, do the following:
F
Check RF system.
F
Perform leak back test.
F
Test throttle valve.
F
Perform vacuum RF test and verify that pressure is stable during the
test.
F
Check capacitance manometer.
F
Replace throttle valve.
PRINTER BUFFER FULL. CALL TECHNICAL
SERVICE.
There is no printed message for this condition.
Problem/Condition
F
The printer buffer has exceeded the maximum allowed value of 23.
F
Cycle completed or canceled with no printout.
Action
Perform a Master Reset.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-27