Align the Chamber
This procedure is not done routinely, it should only be done if needed as
part of a door replacement, or if the chamber is out of alignment for
another reason.
1. Loosen the brackets and gently move the chamber from side to side
to bring it into alignment.
2. On Block 2, loosen the bracket at the side to move the chamber.
Figure 7-5. Cable Adjustment
Door and Chamber Subsystem
Calibration
7-12
STERRAD® 100 Sterilizer
ASP Service Guide rev D