POSITIONING CASSETTE
There is no printed message for this condition.
Problem/Condition
The cassette is being positioned. The reflective sensor did not detect the
side cell positioning mark. The cassette advances to the next position.
Action
F
Inspect the cassette for damage or smudges on bar-code.
F
Insert a new cassette.
F
Test the pump optics and electronics
Power interrupted.
Problem/Condition
Possible brief power failure.
Action
Check customer log to determine if there is an ongoing problem in this
area. If this is a chronic problem, site may have chronic power problems
that need to be addressed by ASP Technical Services and customer’s
engineering department.
Power interrupted. Call ASP Technical
Service.
Problem/Condition
Power was interrupted for the second time and for more than 1 minute.
Action
F
Check incoming power supply. Verify that phase is correct.
F
Check for loose connections.
F
Replace the AC enclosure