Align Chamber To Left Guide Rail
1. Assure that the left guide rail is flush with the frame front edge and
secured with mounting hardware. Adjust if necessary.
2. Remove the screw and cable from the left upper door stop bracket to
give clearance for gage T.2266.
3. Using gage T.2266 align the chamber to the left guide rail. See the
diagram. Adjust the two horizontal mounting slots, located at the
rear of the chamber. This sets the chamber face parallel to left guide
rail, giving a slip fit.
4. Tighten the three bolts securing the left guide rail to chamber.
5. Using gage T.2266 verify that the alignment of the chamber and the
left guide rail has not changed.
Figure 7-2. Align chamber to Right Guide Rail
Calibration
Door and Chamber Subsystem
ASP Service Guide rev D
STERRAD® 100 Sterilizer
7-9