Low plasma power. Repackage load.
Restart sterilizer.
Problem/Condition
After 2 minutes, RF delivered power is less than 365 W but greater than
200 W.
Action
F
Ensure the shelves are not in contact with the chamber or the door.
F
Make sure the load is not in contact with the electrode.
F
Repackage the load.
F
Restart the sterilizer.
Low RF Power in Vacuum. Call ASP
Technical Service
Problem/Condition
RF delivered power is less than 365 W but greater than 200 W.
Action
F
Remove the shelves and run an empty chamber cycle. If completed
correctly, reinstall the shelves.
F
Check the RF subsystem
Low RF Power. Call ASP Technical
Service
Problem/Condition
RF delivered power is less than 365 W but greater than 200 W.
Action
F
Remove the shelves and run an empty chamber cycle. If completed
correctly, reinstall the shelves.
F
Check the RF subsystem for malfunction.
System Messages
Using System Messages
3-14
STERRAD® 100 Sterilizer
ASP Service Guide rev D