If the pressure exceeds the set limits, the cycle is canceled with the
message “Pressure Out of Range.”
The RF power is delivered to the chamber at 400 +25 -35 Watts. If the
RF Power is not within the set limits, the cycle is canceled with one of
the messages “RF Amplifier Failure,” “Plasma Not Lit,” and “Low Plasma
Power.”
Vent Stage
At the end of the Plasma Stage, the RF energy is turned off and the
process chamber is vented to the atmosphere through a bacteriological
(HEPA) filter. The chamber is pumped down again to 1 Torr and then
again vented to the atmosphere.
At the end of the second vent stage, which is the end of the sterilization
cycle, an audible alarm tells the operator that the door may be opened
and the sterilized contents removed from the process chamber. The
printer prints a summary of cycle parameters, and the LCD displays
“Process Complete.”
First Vent
Duration
Parameter
Limit
Variable
Pressure switch
~690 Torr
Pump Down
Duration
Parameter
Limit
Variable
Pressure
1 Torr
Final Vent
Duration
Parameter
Limit
Variable
Pressure switch
~690 Torr
Theory of Operation-Overview
System Operation
ASP Service Guide rev. D
STERRAD® 100 Sterilizer
2-9
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......