Injection system interrupted.
Problem/Condition
Bad cell in cassette.
Action
Check the customer log to determine if there is an ongoing problem in
this area.
Injection System Interrupted See Op
Manual Appendix A
Problem/Condition
The sterilization process canceled because the cassette was faulty.
Action
F
The faulty cassette ejects into the collection box.
F
Insert a new cassette.
F
Repackage the load.
F
Restart the sterilizer.
INSERT NEW CASSETTE
There is no printed message for this condition.
Problem/Condition
There is no cassette in the machine.
Action
Insert a new cassette.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-9