CYCLE 2 LONG CYCLE
YES(START)—NO(CANCEL)
F
If you are processing a flexible scope, press
START
to start a long
cycle.
F
If you are not processing a flexible scope, press
CANCEL
to go to the
CYCLE 1 (
short cycle option.
Door Going Up and Down
Problem/Condition
Equipment has been serviced and now door doesn’t work.
Action
F
Check connector on top of optics board. It may have been knocked
loose during servicing.
F
Check for foreign objects under the counterweight.
F
Check door sensors.
F
Perform door test.
Door Stuck in Closed Position
Problem/Condition
Door sensors pass tests but door cannot be opened by either throwing
the circuit breaker or using AutoTest. Door solenoid light is lit (red).
Pressure gauge is in range (60-80 PSI) and compressor is operational.
There is NO vacuum in the chamber.
Action
F
Take the pneumatic door cylinder pressure hoses and switch them.
(One hose blows air.) Door should open but cannot now be closed.
F
Replace door pneumatics solenoid.
F
Check vacuum switch.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-5
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......