WARMING UP
Problem/Condition
Chamber is not a operating temperature.
Action
Allow chamber to warm up before attempting to run a cycle.
WARMING UP CYCLE WILL START
Problem/Condition
Chamber is not at operating temperature.
Action
F
Cycle starts automatically.
F
Press
CANCEL
to remove load before start.
WARMING UP PLEASE CLOSE DOOR
Problem/Condition
Chamber is not at operating temperature.
Action
Close the door and allow chamber to warm up before attempting to run a
cycle.
Suspected Leak
Problem/Condition
The leak back test results are greater than 25 mTorr; or there is evidence
of a large leak in that the system cannot pump down below 700 mTorr.
Action
F
Identify source of leak.
F
Replace the failed unit.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-43