F
Also, verify voltage at P95 in the door reads 1 VAC.
Corrective Action
Replace solid state relay board.
Display
HEAT ON W/ START
NEXT OUTPUT W/ CANCEL
Action
Press
CANCEL
for next output.
Oil Return Valve Tests
Display
4
4
Note:
The following test is for Block 2 and all STERRAD 100 S
machines. On Block 2 there is a 120 VAC on the
connector. On Block 1.8 (with STERRAD 100 S software)
there is a 12 VDC connector.
ON/OFF OIL SW W/ START
NEXT OUTPUT W/ CANCEL
Action
Press
START.
Display
OIL SWITCH ON W/ START
NEXT OUTPUT W/ CANCEL
Action
Verify the oil switch is on by measuring the voltage connector on the oil
return valve or, listen to hear the solenoid click
Corrective Action
F
Check wiring.
F
Replace solid state relay board.
F
Replace digital I/O board.
F
Replace switch.
Throttle Valve/Heater Menus