Setting the Audible Alarm
Block 2 systems allow for setting the volume of the audible alarm. The
switch is on membrane control PCB at the back of the membrane switch
panel.
1. Turn the bottom blue switch (R4) clockwise to increase volume and
counter-clockwise to decrease.
Setting the LCD Contrast
All systems allow your to set the LCD display contrast. The switch is on
the membrane control PCB at the back of the membrane switch panel.
1. Turn the top blue switch (R1) clockwise to increase volume and
counter-clockwise to decrease.
Setting System Parameters
System parameters or defaults are those settings that keep track of
system activities. System activities are regularly included in the
messages printed by the system. They are:
F
STERRAD 100 S System Settings
F
Setting the System Serial Number and Total Machine Cycles
F
Setting the System Clock
F
Master Reset
STERRAD 100 S System Settings
Caution: If power is off for more than 1 minute, the system
goes to the Cancel/Abort processing as soon as
power is restored, and the printout indicates that
the cycle did not complete properly. If power is off
for less than 1 minute, however, the door closes and
the cycle continues when power is restored; the
printout does not indicate that there is a problem.
Results from any cycle in which the door opened
during the sterilization process should be considered
invalid, and the load should not be considered
sterile.
User Interface and Control Subsystems
User Interface Calibration
6-12
STERRAD® 100 Sterilizer
ASP Service Guide rev D.
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......