DOOR IS BEING BLOCKED
There is no printed message for this condition.
Problem/Condition
Something may be blocking the door or the over-center sensor was
activated before the top sensor.
Action
Check the door top sensor, the over center sensor, the door position, the
over center flag and the harness connection.
DOOR WILL CLOSE AUTOMATICALLY
There is no printed message for this condition.
Problem/Condition
Indication that the door will close normally because
START
has been
pressed.
Action
No action needed.
EJECT TIME OUT CALL TECHNICAL SERVICE
There is no printed message for this condition.
Problem/Condition
Cassette ejection is greater than 20 seconds or cassette capture is more
than 30 seconds.
Action
F
A cassette is stuck in the injector pump. Remove it.
F
Check injector pump optics
F
Check injector pump pneumatic solenoid.
INCORRECT CASSETTE TYPE