Corrective Action
F
Check throttle valve harness connections
F
Replace throttle valve.
F
Perform a Leak Back Test.
RF Test Set-Up and Test
RF Test Set-Up
WARNING:- RF power is present during this operation. do not
touch the RF generator or any of the wiring.
F
If peroxide has been injected, run the RF for 15 minutes. Monitor
the reflected power display in mV. The value begins as high as 500
mV then is decreases to a value typically below 100 mV. If the value
begins to climb above 300 mV, there is a potential mismatch in
impedance.
F
After 500 mTorr is reached and is stable, press a
LONG
CANCEL
for
the RF Test menu.
F
Set-up a Bird Watt meter on the X10 scale. Reading is on the
Kilowatt scale.
F
Connect the RF generator to the INPUT of the directional coupler on
the Watt.
F
Connect the RF interconnect cable to the OUTPUT of the directional
coupler Watt. The RF element arrow should point towards the RF
match.
RF Test
Display
P= 500 mT,
TURN RF ON W/ START
Action
F
Press
START
.
F
Verify operation of RF by observing RF power measurement using the
Bird Watt meter. Observe for 15 minutes.
F
Clean the electrode, RF feedthrough and chamber using deionized
water and a Scotch Brite-type scrub pad.
AutoTest
Vacuum Subsystem Tests
B-36
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......