Display
CHAN #7 HIGH PRESS CHNL
CONTINUE W/ START
Display
#.### V. HIGH PRESSURE
NEXT INPUT W/ CANCEL
Action
F
This pressure reading should be exactly half the channel 0 reading.
If not, adjust pot R7 (Block 2 is R5) on the connector distribution
board until the LCD reads exactly half the channel 0 reading.
F
Apply a small amount of Glyptal to R7 (Block 2 is R5) to lock in
place after adjustment is competed.
F
Remove adjustment tool from pressure sensor connector and
re-connect the harness to Baratron pressure sensor.
F
Press
CANCEL.
Corrective Action
F
Verify that the adjustment tool’s impedance is correct.
F
Replace the control enclosure if the adjustments cannot be made.
Display
EXIT ANALOG INPUT
WITH LONG CLOSE
Action
Press a
LONG
CLOSE.
Display
ANALOG W/ START
NEXT Menu W/ LONG CANCEL
Action
F
To restart test, press START, otherwise press a
LONG
CANCEL.
AutoTest
Analog Test
B-32
STERRAD® 100 Sterilizer
ASP Service Guide rev D