
Display
MOTOR OUT W/ START
NEXT OUTPUT W/ CANCEL
Action
F
Press
START
or
CANCEL.
Injection Test
Display
ON/OFF INJECT W/ START
NEXT OUTPUT W/ CANCEL
Action
WARNING! Exposure to H2O2! DO NOT use a cassette with ANY
full cells for the following test.
Verify injection
F
Insert a used, empty cassette into the injector pump and verify that
it glides easily between the guides without pinching (0.5 mm). ·
F
Ensure that the pawl is fully engaged in the cassette.
F
Press
START
to extend the cam and verify that the roller sits on the
flat of the cam when fully extended.
F
Press
START
to retract cam. When the cam is retracted, remove the
cassette.
Verify Valve Height
4
4
Note
: Block 2 procedures are shown below.
1. Press START to fully extend the cam and insert the injector cassette
gap tool from the back of the platen.
2. Push the platen down to allow the tool to advance on top of injector
valve.
3. Verify that seal block o-ring and needle seals clear the opening in the
cassette gap tool.
4. Turn the injector valve mounting screws to raise the valve until no
gap exists between seal block and tool or between the needle seal
retainer plate and tool.
5. Remove the cassette gap tool.
Injection and Pneumatics Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-17
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......