5. Attach the labeled cover to the rear of the AC Enclosure and secure
by hooking the tabs into the slots and securing the PEM studs inside
the enclosure with 2 each, M4 Flat washers and M4 Hex Lock nuts.
Thermal Overload Relay
1. Attach the thermal overload relay to the AC interface enclosure.
2. Switch the mounting pins so that the two movable pins are set to
outboard (widest) location.
3. Lift the clear plastic cover and set the overload to 3.6 amps, for the
low voltage units (200 VAC, 208 VAC) and 1.8 amps for high voltage
units (308 VAC, 398 VAC, 415 VAC).
4. Replace the clear plastic cover.
4
4
Note:
Terminations ending in a clamping device should have
only the stripped wire inside the clamp. The insulation
should not be inside the clamp.
5. Connect orange wire #80 to CR3-NC 95.
6. Connect orange wire #90 to CR3-NC 96.
7. Connect black wire #29 to CR3-T1.
8. Connect blue wire #30 to CR3-T2.
9. Connect red wire #31 to CR3-T3.
10. Install CR3 to CR2 by hooking CR3 into the cover of CR2 and
inserting pins into CR2-T1, T2 and T3. Secure the terminal screws.
11. Attach the configuration board to the standoffs in the control
enclosure and plug the connectors into the board.
International System Configurations
System Installation
ASP Service Guide rev. D
STERRAD® 100 Sterilizer
11-17
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......