2. Connect a DVM between the ground (terminal #1) which is closest to
the connector distribution board edge, and the +12 Volts, Terminal
#3. On Block 2 devices connect a DVM between the ground (pin #8
at the top of J121) and +12V (pin #3).
3. Adjust the 12 volt power supply to 12.85 volts, 0.05 volts with the
voltage adjust trim pot located in the lower center area of the 12 volt
power supply board. On Block 2 devices the trim pot is located in the
lower right area of the 12 V power supply board.
Figure 6-4. Block 2 Power Supply Adjustments
Figure 6-5. Block 1 Power Supply Adjustments
User Interface and Control Subsystems
Controller Calibration
6-20
STERRAD® 100 Sterilizer
ASP Service Guide rev D.