CASSETTE INSERTED INCORRECTLY
There is no printed message for this condition.
Problem/Condition
The bar-code cannot be read. The cassette ejects.
Action
F
Insert a new cassette making sure it is inserted in the correct
orientation.
F
If the problem persists, inspect the cassette, check the injector pump
optics.
F
Repair and recalibrate if necessary.
Cassette is Not Accepted
Problem/Condition
System appears to be operating normally following adjustment or PM,
however when trying to run a cycle, the unit does not accept the cassette.
Action
F
If you have been servicing the injector valve you may have failed to
replace the pneumatic tubing. Reconnect it to the injector pump.
F
Perform cassette optics test.
CASSETTE OUT-OF-DATE
There is no printed message for this condition.
Problem/Condition
This cassette has expired or the sterilizer system date is incorrect.
Action
F
The cassette automatically exits to the collector box. Insert a new
cassette.
F
Check system date—pay particular attention to the year. Reset if
incorrect.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-3
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......