Vacuum-RF Test
Display
DOOR CLOSE/OPEN W/ START
NEXT W/ LONG CANCEL
Action
Press
START.
Display
CLOSE DOOR W/ START
NEXT W/ LONG CANCEL
Action
F
Press
START
.
F
Verify operation of door.
F
Verify that chamber O-ring is seated against door face.
Corrective Action
Realign door.
Display
DOOR CLOSE, NEXT VENT
NEXT W/ LONG CANCEL
Action
Press a
LONG CANCEL.
Display
CLOSE VENT W/ START
Action
F
Press
START
.
F
Verify vent is closed.
AutoTest
Vacuum Subsystem Tests
B-34
STERRAD® 100 Sterilizer
ASP Service Guide rev D