4
4
Note:
Prior to tightening the screws on the front panels, verify
the alignment of the cassette through the cassette door
by inserting a cassette.
7. Reinstall the front panel assembly and replace the screws fastening
the front panel to the frame assembly. The screws are reached from
the left and right side of the system at the rear of the front panel
assembly. One set is located at the top of the panel, one set at the
middle, and one set at the bottom of the panel. On Block 2 devices
reattach the front panel and re-install the screws on the brackets.
8. Reconnect the cassette door assembly cable. Close the chamber
door. Close the storage drawer.
Remove/Replace the Door Seal O-Ring
1. Using a plastic tool at the wide hole in the O-ring groove, lift the
O-ring from the groove. Discard the O-ring according to facility
procedures.
2. Clean the groove thoroughly, using alcohol and a lint free cloth.
3. If the new O-ring is not lubricated, thoroughly, but lightly coat it with
a vacuum grease recommended for high vacuum applications.
4. Starting at the top of the groove, press the O-ring into the groove at
two or three inch intervals until it is held in place by the groove.
5. Then press it down completely into the groove making sure it is fully
seated. Be careful not to stretch the O-ring as you work.
Remove/Replace the Door
Open The Door For Servicing
1. Open the storage drawer to reveal the right guard rail. Reach behind
the open drawer and find the door stop lever.
2. Press
CLOSE
DOOR
. When the door moves up, set the door stop
lever down and press
OPEN
DOOR
. The door opens to its service
position, exposing the door O-ring that seals the chamber door, the
rear surface of the door assembly and the chamber. See the diagram
Remove the Door
This procedure is for removing a standard door and installing a standard
door. Information on installing a “show” door is included in the
replacement section.
4
4
Note:
One step in this procedure requires the assistance of
another person.
1. Remove the system panels. Be sure to unplug the cassette door
harness.
Removal/Replacement
Door and Chamber Subsystem
ASP Service Guide rev D
STERRAD® 100 Sterilizer
7-15
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......