Corrective Action
F
Replace solid state relay board.
F
Replace vent valve.
Display
VENT CLOSE, PUMP’S NEXT
NEXT W/ LONG CANCEL
Action
Press a
LONG CANCEL.
Display
PUMP ON/OFF W/ START
NEXT W/ LONG CANCEL
Action
Verify that throttle valve is in open position, indicated by a green LED on
the throttle valve board, and press START.
Display
PRESS 15 TORR, WAIT 600mT
NEXT W/ LONG CANCEL
Action
F
As vacuum reaches 600 mTorr, observe the LED on the throttle
valve. When the pressure reaches 525 mTorr, the throttle valve
should close. (Indicated by a red LED on the throttle valve board.)·
F
After the throttle valve is closed, manually inject H2O2 by pressing
and holding the blue button (Block 2) on the front solenoid valve at
the pneumatic control assembly or the orange button (Block 1.8) on
the right solenoid valve.
F
Observe the pressure reading on the control panel, it should increase
rapidly. Release the button when the pressure reading stops
increasing.
F
After 500 mTorr is reached and is stable, press a LONG CANCEL for
the next menu.
Vacuum Subsystem Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-35
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......