F
The software calculates the midpoint of the minW numbers obtained
in the previous steps.
F
Adjust the fiber optic amplifier trim pot counterclockwise 1 turn and
reinsert the cassette.
4
4
Note:
The cassette should continue to cycle in and out. If it
does not, continue to adjust the side slot trim pot 1/4
turn COUNTER-CLOCKWISE until the injector pump
accepts the cassette.
F
While the cassette continues to cycle in and out, adjust fiber optic
amplifier trim pot to obtain the midpoint minW number calculated in
the previous step.
F
Re-verify that the speed (SL AVR) is still 500 ±5 at this setting. Ignore
the values for BC minB and SL minB.
F
Press
START
. The system runs a 10 cycle verification.
Corrective Action
F
If adjustments cannot be achieved, replace injector pump board.
F
Exit by pulling cassette out of injector pump from the front or the
rear, whichever is easier.
Display
INSERT NEW CASSETTE
Action
Press a
LONG CLOSE.
Display
CASSETTE ADJ W/ START or
NEXT Menu W/ LONG CANCEL
Action
Press a
LONG CANCEL.
4
4
Note:
The following test is not performed at this time.
Display
SLOT ADJ. W/ START
NEXT Menu W/ LONG CANCEL
Cassette Adjustments
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-41
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......