Thermocouples
Technical Data
ASP Service Guide rev. D
STERRAD
®
100 Sterilizer
A-10
Thermocouples
Thermocouples
Measurement
Location
TC1
42°-50° C
Top of chamber behind injector valve
TC2
42°-50° C
Bottom of chamber near door
TC3
42°-50° C
Middle of door
TC4
42°-50° C
Edge of door
Tools and Test Equipment
The following items may be needed to perform the service procedures for
the STERRAD 100 Sterilizer. Included are parts needed for chamber
heater and feedthrough replacement.
Manufacturer
Model or ASP Part
Number
Where Used
Metric Tool kit
Local Supplier
All maintenance
Digital multimeter
(DMM) accurate to
3 decimal places
Fluke 87 or equivalent
Product certification and calibration
Temp. Probe -50`C to
+150`C
Fluke 80T-150U
Product certification and calibration
RF Meter with 50 Ohm
impedance
Bird 4412
Product certification
Service Test PWA
04-03295-0-001
Repairs product certification and
calibration
Autotest PWA (available
to ASP Service
Personnel only)
04-02397-0-001
Repairs product certification and
calibration
Adjustment Tool
Baratron
35-02211-001
Product certification and calibration
Ball driver
35-02716-011
Planned maintenance
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......