Action
F
Value may vary.
F
Press
CANCEL.
Display
CHAN #4 FLOATING INPUT
CONTINUE W/ START
Action
Press
START.
Display
4.095V., FLOATING POINT
NEXT INPUT W/ CANCEL
Action
Press
CANCEL.
This channel is currently unused.
Display
CHAN #5 DOOR TEMPERATURE
CONTINUE W/ START
Display
#.### V. DOOR TEMP.
NEXT INPUT W/ CANCEL
Display
CHAN #6 CHAMBER TEMP
CONTINUE W/ START
Display
#.### V. CHAMBER TEMP.
NEXT INPUT W/ CANCEL
Analog Test
AutoTest
ASP Service Guide rev D