Technical Data
Environmental Conditions
A-5
STERRAD
®
100 Sterilizer
ASP Service Guide rev D
Installation Space
Requirements
In operation, the system shall not be placed
closer than 50 mm (2 in) to a wall at the rear
and/or sides of the system. The system
should be installed in a space of sufficient
size to permit access to all four sides of the
system when the system is rolled away from
the rear and/or side walls a distance of 3
meters (10 ft), the extent of the power cord.
Service access should allow for a minimum
clearance of one meter (3 ft) on all sides of
the system.
Mobility
The sterilizer is designed to be free standing
and self contained with no required
connections other than the electrical power
cord. The sterilizer is equipped with wheels
that allow it to be rolled easily from one
location to another. The unit is also
equipped with adjustable feet at the front of
the unit if leveling is required.
Electrical Requirements
Voltage Specification
IMPORTANT
WARNING!
Prior to relocating the sterilizer to a new power
source, electrical phase rotation should be
checked by a qualified technician. Failure to
verify phase rotation may cause damage to the
system and void the warranty.
208V 60 Hz
Power200V and
208V (
Block 2
)
For versions employing 208V, 60 Hz power,
the sterilizer requires a NEMA L21-20 five wire
grounding twist lock outlet attached to a
dedicated 20 Amp 3 phase 208 Volt circuit
with separate neutral and ground conductors.
380 - 415V 50 Hz
Power380V, 398V
and 415V (
Block 2
)
The sterilizer requires a five wire grounding
outlet attached to a dedicated 10 AMP, 3
phase, 380V circuit with separate neutral and
ground conductors.
Phase Rotation
The system requires CBA phase rotation. The
phase rotation is adjusted to match the
system requirements at installation.
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......