F
When the door is closing the LCD displays.
DOOR IS CLOSING
TEST CENTER, BLOCKED DOOR
F
While the door is closing, press down on the protective block to verify
retracting.
Corrective Action
F
Check the 12 V power supply.
F
Adjust or replace the door over-center sensor.
F
Replace the door solenoid.
Display
CLOSE THE DOOR W/ START or
EXIT DOORTST, LONG CLOSE
Action
Press
START
.
When the door is closing the LCD displays:
DOOR IS CLOSING
TEST CENTER, BLOCKED DOOR
When the door is closed the LCD displays:
TOP AND CENTER BLOCKED
OPEN WITH START
Action
F
Press
START
.
F
When the door opens the LCD displays:
CHECK BOTTOM SENSOR
IS NOT BLOCKED
Verify that the bottom sensor is not blocked.
Corrective Action
Adjust flag or replace bottom sensor.
AutoTest
Door Test
B-28
STERRAD® 100 Sterilizer
ASP Service Guide rev D