Low pressure in injection.
Problem/Condition
Possible bad cell in cassette or partially clogged vaporizer cap.
Action
Check customer log to determine if there is an ongoing problem in this
area. Additional customer training may be necessary.
Low pressure in injection. Call ASP
Technical Service.
Problem/Condition
F
For the second consecutive cycle, the pressure did not rise above 6
Torr at the end of the injection stage.
F
After adjustments the highest reading obtainable is approximately
5.35 Torr.
Action
F
Clean the vaporizer plate.
F
Clean the injector valve vaporizer bowl or cap.
F
Run an empty chamber cycle. Injection pressure must be above 8.5
Torr.
F
Check for peroxide running down the sides of the injector valve and
interfering with heater contact.
F
Make sure the heater is plugged in and voltage is 24 VAC.
F
Inspect the injection subsystem for malfunction including possible
dissolved outer coating due to excessive hydrogen peroxide contact.
F
Replace the injector valve.
System Messages
Using System Messages
3-12
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......