Remove the AC Enclosure
1. Remove the system panels. Unplug all subsystems: vacuum pump,
compressor.
2. (Block 2) Disconnect the vacuum pump harness and grounding strap.
3. Unscrew the unit from the frame. All the screws are on one side of
the unit.
4. Remove the front panel and set it aside. It must be attached to the
new enclosure.
5. (Block 2) Remove the thermal overload relay and the configuration
board from the old enclosure and set them aside. Be sure to place the
configuration board in a static safe area.
Figure 9-2. AC Enclosure Block 2
AC Subsystem
AC Removal/Replacement
9-4
STERRAD® 100 Sterilizer
ASP Service Guide rev. D