Continuous Positive BIs
Problem/Condition
The biological Indicator test is repeatedly positive. The sterilizer is
running normally in all other respects and the operator is performing the
tests correctly.
Action
F
Ensure that vaporizer cap is free from damage and does not show
signs of abnormal wear. Clean and or replace as necessary.
F
Check injector valve height.
F
Check cassette associated with the suspect process cycle. Open
cassette and inspect for damage, residual H2O2, or physical
abnormality.
F
Review process parameter printer tape for suspect process cycle.
Verify recorded injection pressure is correct.
F
Review cycle history for previous 30 days. Report any process
parameter printouts and records for any previous positive or suspect
positive cycles.
F
Make sure electrode screw is tight.
F
Make sure all RF system connectors and screws are tight.
F
Replace the electrode.
System Messages
Using System Messages
3-44
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......