19. If you are going to move the unit to a different site, replace the side
and lower back panels.
20. Once the unit is in position in the installation site, remove all exterior
panels and temporarily store them in a safe location.
Figure 11-2. Positioning Diagram
System Installation
STERRAD® 100 Sterilizer
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......