Display
EXIT SENSOR TEST
W/ LONG CLOSE
Action
Press a
LONG
CLOSE.
Display
SENSOR TEST W/ START or
NEXT Menu W/ LONG CANCEL
Action
To repeat test press
START
, otherwise press a
LONG
CANCEL.
Injection and Pneumatics Tests
Display
INJPNEU W/ START
NEXT Menu W/ LONG CANCEL
Action
F
Press START.
Pawl Test
Display
ON/OFF PAWL W/ START
NEXT OUTPUT W/ CANCEL
Action
F
Press
START
. Verify that the pawl has lifted.
F
Insert a cassette and verify that the pawl clears the cassette by 0.2
mm minimum (.008). Remove the cassette.
AutoTest
Injection and Pneumatics Tests
B-14
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Summary of Contents for STERRAD 100