M
750 mTorr
Max. pressure rise in 10 min.
(38-48 min.)
C
525 mTorr
Plasma high pressure alarm
C
475 mTorr
Plasma low pressure alarm
C
690 Torr
Vent (atmospheric pressure switch)
C
1 Torr
Vent pump-down threshold
RF Delivered Power
Directional coupler in RF amplifier board.
Control/
Monitor
Alarm Limits
Process Equipment
Specification
M
2.92 Volts
Minimum allowable power (365 W)
Temperature
Thermistors are located on the chamber and door to monitor
temperature.
Control/
Monitor
Alarm Limits
Process Equipment
Specification
C
40°C
Lower limit
C
50°C
Upper limit
M
60°C
Temperature alarm threshold
Software Overview
Software controls the STERRAD® 100 Sterilizer to ensure the most
efficient sterilization process as well as ease of use and operator safety.
The software manages the sterilization process throughout the five
stages: vacuum, injection, diffusion, plasma, and vent. The software also
monitors pressure, temperature, and other factors to determine when to
move from one stage to another, what to adjust in the various stages,
and when to abort the process if there are problems. The software
controls the machine in the monitor state including the processing of
hydrogen peroxide cassettes.
Theory of Operation-Overview
System Operation
ASP Service Guide rev. D
STERRAD® 100 Sterilizer
2-13
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......