Remove the Vent Valve
5. Using two wrenches, unfasten the VCO connector securing the vent
valve assembly. Remove by hand taking care not to lose or damage
the O-ring inside the vacuum switch.
Remove the HEPA Filter
6. Unscrew the HEPA filter element from the nipple attached to the vent
valve. Re-tape the nipple with Teflon tape prior to replacement.
Remove the Intake Filter
7. Remove the throttle valve first. Unfasten the KF clamps securing the
intake filter assembly to the throttle valve and the pump. Remove the
centering ring and lower elbow if applicable.
Remove the Exhaust Filter
8. Disconnect the oil return hose (Block 2).
9. Unfasten the KF-25 clamp securing the exhaust filter to the pump.
Remove the AC Connection
10. Unfasten the ground from the cabinet (Block 2) or the ground
connection on the AC harness adapter. Unplug the pump from the
AC enclosure (Block 2.0) or from the AC harness adapter. Cut the
cable ties and free the cable.
Remove the Drain Valve
11. Unscrew the drain valve as an entire unit.
4
4
Note:
Drain valve replacement requires the use of the drain
valve kit.
Vacuum Subsystem
Removal/Replacement Procedures
4-10
STERRAD® 100 Sterilizer
ASP Service Guide rev. D
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......