8. Use
OPEN
DOOR
and
CLOSE
DOOR
in the same way as shown in
steps 5 and 6 to set the serial number.
9. When you have set the serial number, press
CLOSE
DOOR
so that
you do not accidentally change the number
10. Simultaneously press
START
and
OPEN
DOOR
(the display switches
to the edit date/time display). Then simultaneously press START and
CANCEL
to return to the active display.
Master Reset
1. Defeat the safety interlock, remove the service test (Autotest) PCB (if
installed), and install the system microprocessor PCB.
2. Apply power to the enclosure by enabling the safety interlock and
pressing CANCEL simultaneously.
3. Continue to hold CANCEL until you hear the beep signal, then
release.
4. Remove the cassette collection box. Wait for the beep tone and then
reinstall the empty box.
User Interface and Control Subsystems
User Interface Calibration
6-18
STERRAD® 100 Sterilizer
ASP Service Guide rev D.