Display
VAPORIZER OFF W/ START
NEXT OUTPUT W/ CANCEL
Action
F
Press
START
.
F
Check voltage at injector valve connector P-46, should be VAC. beep
warning should stop.
Corrective Action
If 5 VAC, or beep sound persists, replace solid state relay board.
Display
VAPORIZER ON W/ START
NEXT OUT PUT W/ CANCEL
Action
Press
CANCEL.
Chamber Door Test
Display
OPEN/CLOSE CHAMBER DOOR
W/ START OR NEXT W/ CANCEL
Action
F
Press
START
.
F
Verify that door sensors are secured and are not subject to damage
during operation. Verify flags are centered in sensors.
F
Adjust the speed control valve on door cylinder.
F
Block 2
. Verify chamber door closes in approximately 5 seconds,
adjust and secure speed control valves on pneumatic control
assembly.
F
Verify that the door is sealed.
F
Verify that door meets chamber door stop.
F
Check that door actuator arms are extended but not over center.
F
Check that guide cables are tensioned.
Injection and Pneumatics Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-19