Display
OIL SWITCH OFF W/ START
NEXT OUTPUT W/ CANCEL
Action
F
Press
START
.
F
Verify the oil switch is OFF by measuring the voltage on the oil
connector.
Corrective Action
F
Check wiring.
F
Replace solid state relay board.
F
Replace digital I/O board.
Display
OIL SWITCH ON W/ START
NEXT OUTPUT W/ CANCEL
Action
F
Press
CANCEL
.
F
Press
LONG
CANCEL
to bypass the next tests.
Cassette Switch Tests
4
4
Note:
The following tests are not applicable for the field. They
are for units with cassette feeders. The displays are
shown for information only.
Display
ON/OFF CASSETTE SWITCH
W/ START OR NEXT W/ CANCEL
4
4
Note:
This signal can only be measured if the machine is
equipped with the harness or tool for the cassette feeder
or if the machine is a Block 2 system.
Display
CASSETTE SW ON W/ START
NEXT OUTPUT W/ CANCEL
AutoTest
Throttle Valve/Heater Menus
B-26
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Summary of Contents for STERRAD 100