Vaporizer out of range Call ASP
Technical Service
Problem/Condition
The current sensor has not detected activation of the vaporizer heater.
Action
F
Check vaporizer heater. Replace if necessary.
F
Replace current sensor.
VENT TIME OUT
Problem/Condition
Atmospheric pressure was not detected by the end of the vent stage in
vacuum, diffusion or at the end of the sterilization cycle.
Action
F
Check time and settings on Whitman Vacuum Switch and reset if
necessary.
F
Check LCD during vacuum switch test. Verify that setting is O. If
not, reset for 2 seconds.
F
Make sure vacuum switch retaining collar is tight.
F
Make sure vacuum switch is operational.
F
Check DC voltages
F
Replace the vacuum switch
F
Replace the pneumatic board.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-41
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......