TEMP. SENSOR FAILURE. NO MORE
CYCLES CAN RUN.
Temperature sensor failure. Call ASP Technical Service
Problem/Condition
Temperature of door or chamber is 0° C.
Action
F
Test the thermistors for possible failure.
F
Perform the heater test.
F
Verify heater voltages
F
Replace thermistor.
UNABLE TO VERIFY INSERT NEW CASSETTE
There is no printed message for this condition.
Problem/Condition
The bar-code cannot be read for the second time. (This message follows
an unsuccessful cassette verification.)
Action
F
Insert new cassette.
F
If the message persists, test the cassette optics.
System Messages
Using System Messages
3-36
STERRAD® 100 Sterilizer
ASP Service Guide rev D