Action
F
Place cassette box on box switch; the beeps should sound.
F
Remove the cassette box from the box switch. No beeps should
sound when the box is lifted.
Action
F
Press
CANCEL
for the next sensor test.
Corrective Action
F
If no beeps, check the wiring and connections.
F
Replace switch.
F
Replace the pneumatic board.
F
Replace the digital I/O board.
Whitman Vacuum Switch Test
4
4
Note:
Not used with Block 2 systems.
Display
TEST WHITMAN VAC SWITCH
W/ START OR NEXT W/ CANCEL
Action
F
Press
START
. Display indicates switch should be disconnected.
F
Disconnect switch to test. Display indicates reconnection.
F
Press
CANCEL.
Corrective Action
F
Adjust the vacuum switch.
SMC Vacuum Switch Test
4
4
Note:
This test is for Block 2.0 systems only.
Display
TEST SMC VAC SWITCH
W/ START OR NEXT W/ CANCEL
AutoTest
Sensor Tests
B-12
STERRAD® 100 Sterilizer
ASP Service Guide rev D