Electrical Requirements
Technical Data
ASP Service Guide rev D
STERRAD
®
100 Sterilizer
A-6
Power Calculation Phase A 228 W-hours per cycle
Phase B 158 W-hours per cycle
Phase C 177 W-hours per cycle
Total 563 W-hours per cycle
Heat Generation
1919 Btu per cycle
Heater Impedance
Heater Location
Type
Impedance
Door
Block 2
190 Ohms
Chamber
Block 2
100 Ohms
Chamber top
Block 1.8
50 Ohms
Chamber bottom
Block 1.8
50 Ohms
Chamber rear
Block 1.8
190 Ohms
Door
Block 1.8
50 Ohms