Single-Ended Signals AC Input Parameters Tests
56
DDR2(+LP) Compliance Testing Methods of Implementation
Probing for Single-Ended Signals AC Input Parameters Tests
When performing the Single-Ended Signals AC Input Parameters tests, the DDR2(+LP) Compliance
Test Application will prompt you to make the proper connections. The connection for the
Single-Ended Signals AC Input Parameters tests may look similar to the following diagram. Refer to
the Connection tab in DDR2 Electrical Performance Compliance application for the exact number of
probe connections.
You can use any of the oscilloscope channels as Pin Under Test (PUT) source channel. You can
identify the channels used for each signal in the Configuration tab of the DDR2(+LP) Compliance
Test Application. (The channels shown in
For more information on the probe amplifiers and differential probe heads, see
Probing,” starting on page 351.
Test Procedure
1 Start the automated test application as described in
“Starting the DDR2(+LP) Compliance
2 Ensure that the RAM reliability test software is running on the computer system where the DDR2
Device Under Test (DUT) is attached. This software will perform a test on all unused RAM on the
system by producing a repetitive burst of read-write data signals to the DDR2 memory.
3 Connect the differential solder-in probe head to the PUTs on the DDR2 devices.
4 Connect the oscilloscope probes to any channels of the oscilloscope.
5 In the DDR2 Test application, click the Set Up tab.
Figure 8
Probing for Single-Ended Signals AC Input Parameters Tests with Two Probes
Infiniium Oscilloscope
InfiniiMax solder-in probes
DDR2 DIMM
Summary of Contents for D9020DDRC
Page 1: ...Keysight D9020DDRC DDR2 LP Compliance Test Application Methods of Implementation ...
Page 10: ...10 DDR2 LP Compliance Testing Methods of Implementation ...
Page 46: ...2 Preparing to Take Measurements 30 DDR2 LP Compliance Testing Methods of Implementation ...
Page 70: ...3 Measurement Clock Tests 54 DDR2 LP Compliance Testing Methods of Implementation ...
Page 224: ...14 Clock Timing CT Tests 208 DDR2 LP Compliance Testing Methods of Implementation ...
Page 270: ...15 Data Strobe Timing DST Tests 254 DDR2 LP Compliance Testing Methods of Implementation ...