DDR2(+LP) Compliance Testing Methods of Implementation
335
Calibrating the Infiniium Oscilloscope and Probe
19
Internal Calibration
This will perform an internal diagnostic and calibration cycle for the oscilloscope. For the Keysight
oscilloscope, this is referred to as Calibration. This Calibration will take about 20 minutes. Perform
the following steps:
1 Set up the oscilloscope with the following steps:
a
Connect the keyboard, mouse, and power cord to the rear of the oscilloscope.
b
Plug in the power cord.
c
Turn on the oscilloscope by pressing the power button located on the lower left of the front
panel.
d
Allow the oscilloscope to warm up at least 30 minutes prior to starting the calibration
procedure in step 3 below.
2 Locate and prepare the accessories that will be required for the internal calibration:
a
Locate the BNC shorting cap.
b
Locate the calibration cable.
c
Locate the two Keysight precision SMA/BNC adapters.
d
Attach one SMA adapter to the other end of the calibration cable - hand tighten snugly.
e
Attach another SMA adapter to the other end of the calibration cable - hand tighten snugly.
3 Referring to
below, perform the following steps:
a
Click on the Utilities>Calibration menu to open the Calibration dialog box.
Figure 41
Accessing the Calibration Menu
Summary of Contents for D9020DDRC
Page 1: ...Keysight D9020DDRC DDR2 LP Compliance Test Application Methods of Implementation ...
Page 10: ...10 DDR2 LP Compliance Testing Methods of Implementation ...
Page 46: ...2 Preparing to Take Measurements 30 DDR2 LP Compliance Testing Methods of Implementation ...
Page 70: ...3 Measurement Clock Tests 54 DDR2 LP Compliance Testing Methods of Implementation ...
Page 224: ...14 Clock Timing CT Tests 208 DDR2 LP Compliance Testing Methods of Implementation ...
Page 270: ...15 Data Strobe Timing DST Tests 254 DDR2 LP Compliance Testing Methods of Implementation ...