16
Data Timing Tests
256
DDR2(+LP) Compliance Testing Methods of Implementation
Probing for Data Timing Tests
When performing the Data Timing tests, the DDR2(+LP) Compliance Test Application will prompt you
to make the propser connections. The connection for Data Timing tests may look similar to the
following diagrams. Refer to the Connection tab in DDR2(+LP) Electrical Performance Compliance
Test application for the exact number of probe connections. Typically, you need minimum three
probe connections to run the tests.
You can use any of the oscilloscope channels as Pin Under Test (PUT) source channel. You can
identify the channels used for each signal in the Configuration tab of the DDR2(+LP) Compliance
Test Application. (The channels shown in
are just examples).
For more information on the probe amplifiers and differential probe heads, see
Probing,” starting on page 351.
Figure 32
Probing for Data Timing Tests with Three Probes
Infiniium Oscilloscope
Infiniium Oscilloscope
DDR2 DIMM
InfiniiMax solder-in probes