DDR2(+LP) Compliance Testing Methods of Implementation
103
Single-Ended Signals VIH/VIL (Address, Control) Tests
5
Measurement Algorithm
1 Sample/acquire signal data.
2 Find all valid positive pulses. A valid positive pulse starts at V
REF
crossing at valid rising edge and
end at V
REF
crossing at the following valid falling edge (See notes on threshold).
3 Zoom in on the first valid positive pulse and perform V
TOP
measurement. Take the V
TOP
measurement results as V
IHCA(AC)
value.
4 Continue the previous step with another nine valid positive pulses.
5 Determine the worst result from the set of V
IHCA(AC)
measured.