49
3
Measurement
3.1
Measurement Workflow
1 Pre-measurement Inspection
Refer to “3.2 Pre-measurement Inspection” (p. 51).
2 Basic Measurement Settings
Refer to...
Select a suitable recording method for the
measurement target
“3.3.1 Measurement Function” (p. 52)
Set sampling rate
“3.3.2 Timebase and Sampling Rate”
Set recording length
“3.3.3 Recording Length (Number of
Set waveform display format
and printing format
Application examples
Refer to...
“7.4 Perform Waveform X-Y Synthesis” (p. 128)
“8.2 Display Waveforms During Recording (Roll Mode)” (p. 147)
“8.3 Display New Waveforms Over Past Waveforms (Overlay)” (p. 148)
“10 Numerical Calculation Functions” (p. 209)
3 Input Channel Settings
Refer to...
Choose analog channel settings
“3.4.2 Choose Analog Channel Settings”
Choose logic channel settings
“3.4.3 Logic Channel Settings” (p. 66)
Application examples
Refer to...
“8.5 Convert Input Values (Scaling Function)” (p. 151)
“8.6 Set the Waveform Position (Variable Function)” (p. 158)
“8.7 Fine Adjustment of Input Values (Vernier Function)” (p. 161)
“8.8 Invert the Waveform (Invert Function)” (p. 162)
4 Trigger Settings
Refer to “9 Trigger Settings” (p. 187).
3 Measurement
3
Measurement
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......