338
External I/O
17.2 External I/O
17.2.1 External Input (START/IN1) (STOP/IN2) (SAVE/IN3)
External control signals can be applied to start and stop measurement and to print and save data. The
factory-default settings are
[Start]
,
[Stop]
, and
[Save]
.
Signal Input Procedure
1
Connect the cables for the corresponding external input signals to the START/IN1, STOP/
IN2, SAVE/.IN3, and GND terminals.
Refer to “17.1 Connect to External Control Terminals” (p. 337).
2
Press the
SYSTEM
key to open the
[Environment]
sheet, and move the cursor to
[START/
EXT. IN1]
,
[STOP/EXT. IN2]
, or
[SAVE/EXT. IN3]
.
3
Select the operation to be performed by the instrument in response to external signal input.
Start
Start measurement. (not affected by
[START Action]
(p. 307))
Stop
Stop measurement. (Operations after measurement, such as numerical
calculations and automatic saving will be carried out)
Start/Stop
Start measurement on LOW level, and stop measurement on HIGH level.
Abort
Stop measurement immediately. (Operations after measurement, such as
numerical calculations and automatic saving will not be carried out)
Same operation as
key (cannot be used with Selective Printing (p. 112))
Save
Save to the media specified for the
SAVE
key, according to the specified
conditions. (cannot be used with Selective Printing (p. 88))
Pen Up/Down
A LOW level signals “pen-down” operation and a HIGH level signals “pen-up”
operation (for X-Y recording).
RUN/STOP
Runs/stops waveform output. (RUN at LOW level, STOP at HIGH level).
PAUSE
Stops waveform output temporarily.
•
For STOP operation, follows the configured
[STOP Action]
• External input is not available when the HELP screen or a dialog window is displayed.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......