238
Operations Available From the [Memory Div] Sheet
• Waveforms can be recorded into individual blocks by dividing memory space into multiple blocks.
• You can record waveforms beginning at any block (
Start Block
), choose which blocks to display (
Display
Block
), or display multiple overlaid blocks (
Reference Block
).
• The maximum number of blocks for memory division depends on the installed memory board and
recording length (up to 1024 divisions).
•
Triggered waveform data can be acquired continuously and recorded sequentially in specified blocks
(from the Start Block, for the specified Used Blocks). Dead time while displaying or printing (during
which triggers are ignored) can be minimized.
• Even if you do not use the Memory Division function, up to 16 blocks of data (depending on the
specified recording length) can be saved to each block so that previously recorded data can be selected
for display on the Waveform Screen.
Refer to “7.8 View Block Waveforms” (p. 137).
Normal
Memory Division
1
2
3
4
5
6
Reference Block
Display Block
The Reference Block can be displayed as an
overlay.
Dead time (not recorded)
Recording length
Recording length
Blocks
When the number of Used Blocks = 5
Start Block
1
2
3
4
5
When batch saving, an index file (SEQ) is created.
When reloading this file, all saved blocks can be
read at once.
Auto Save
Auto Print
Used Blocks
When saving manually, select whether to save all used blocks or just the
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......