Module Specifications
18.6.3
Model 8968 High Resolution Unit
Accuracy figures assume installation in a Memory HiCorder and operation after a 30-minute warm-up
period at 23°C ±5°C (73°F ±9°F) and 80% RH (no condensation).
Product warranty
period
3 years
Guaranteed accuracy
period
1 year
Number of input
channels
2 channels
Measurement range
5, 10, 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20 V/div
Measurement
accuracy
±0.3% f.s. (Filter 5 Hz On, after zero adjustment)
Temperature
characteristic
±0.045% f.s./°C
Frequency
characteristic
DC Coupling: DC to 100 kHz −3 dB
AC Coupling: 7 Hz to 100 kHz −3 dB (low cut-off frequency: 7 Hz±50%)
Noise
500 μV p-p (typ.), 1 mV p-p (max.) (highest sensitivity range, with input shorted)
Common mode
rejection ratio
80 dB or more (at 50 Hz/60 Hz and with signal source resistance 100
Ω
or less)
Low-pass filter
Off, 5±50%, 50±50%, 500±50%, 5 k±50%, 50 k±50% (Hz) −3 dB
Anti-aliasing filters
Cutoff frequency (fc) 20, 40, 80, 200, 400, 800, 2 k, 4 k, 8 k, 20 k, 40 k (Hz)
(Automatic setting when the anti-aliasing filter is ON)
Attenuation property: −66 dB or higher at 1.5 fc
Input type
Unbalanced input (floating)
Input coupling
AC/DC/GND
Input resistance
1 M
Ω
±1%
Input capacitance
30 pF±10 pF (at 100 kHz)
A/D resolution
16 bits
Maximum sampling
rate
1 MS/s
Input terminals
Insulated BNC terminal
Maximum input
voltage
400 V DC
Maximum rated
voltage to earth
300 V AC, DC (between individual input channels and enclosure as well as between
individual input channels)
Measurement category II (anticipated transient overvoltage: 2500 V)
Operating
temperature and
humidity
As per Memory HiCorder in which Model 8968 is installed
Operating
environment
As per Memory HiCorder in which Model 8968 is installed
Storage temperature
and humidity
-10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation)
Dimensions
Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass
Approx. 250 g (8.8 oz.)
Effect of radiated
radio-frequency
electromagnetic field
±15% f.s. (max.) at 3 V/m
Specifications
18
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......