243
13 FFT Function
13
FFT Function
13.1 Overview and Features
FFT analysis can only be used with the FFT function.
The FFT (Fast-Fourier Transform) functions perform frequency analysis on input signal data.
Use these functions for frequency analysis of measurement targets such as rotating objects, vibrations,
and sounds.
For details, refer to “Appx. 4 FFT Definitions” (p. A13).
Analysis can be performed on data as it is being measured, on pre-existing analog waveform data
previously acquired with the Memory function, and on data output from waveform calculations.
When using a Model 8968 High Resolution Unit equipped with an
anti-aliasing filter, the cut-off frequency
can be set automatically by linking with the frequency range setting.
Major Features
• FFT analysis frequency range: 133 mHz to 8 MHz
• FFT Analysis Modes (16 types)
•Storage Waveform
•Cross Power Spectrum
•Histogram
•Impulse Response
•Linear Spectrum
•Coherence Function
•RMS Spectrum
•Phase spectrum
•Power Spectrum
•Auto-correlation Function
•Power Spectrum density*
•Cross-correlation Function
•LPC analysis (Power Spectrum
Density)*
•1/1 Octave Analysis*
•Transfer Function
•1/3 Octave Analysis*
* Not available when using external sampling.
For phase spectra, only the required phase information is highlighted and displayed.
Refer to “13.3.8 Emphasize Analysis Results (Phase Spectra Only)” (p. 255).
When performing FFT analysis with the instrument connected to a sound level or vibration meter,
scaling by dB can be set from the Channel Settings screen if you want to read values directly in
calibrated units of measurement.
To suppress the effects of aliasing distortion caused by sampling
We recommend using Model 8968 High Resolution Units that are equipped with anti-aliasing filtering to
suppress the effects of aliasing distortion when sampling.
Refer to “Appx. 4 FFT Definitions” (p. A13).
Refer to “18.3.4 FFT Function” (p. 352) for FFT function specifications.
13
FFT
Function
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......