366
Module Specifications
Power frequency mode
Measurement
range
Measurement
accuracy
50 Hz (40 Hz to 60 Hz), 60 Hz (50 Hz to 70 Hz), 400 Hz (390 Hz to 410 Hz) (f.s. = 20 div)
±0.03 Hz (50 Hz, 60 Hz), ±0.1 Hz (400 Hz)
Count mode
Measurement
range
Measurement
accuracy
Measurable range
2 k, 10 k, 20 k, 100 k, 200 k, 1 M counts/div
±range/2000
DC to 100 kHz (minimum pulse width: 2 μs)
Duty ratio mode
Measurement
range
Measurement
accuracy
Measurable range
5%/div (f.s.= 20 div)
±1% (10 Hz to 10 kHz)
±4% (10 kHz to 100 kHz)
10 Hz to 100 kHz (minimum pulse width: 2 μs)
Pulse width mode
Measurement
range
Measurement
accuracy
Measurable range
500 μ, 1 m, 5 m, 10 m, 50 m, 100 ms/div (f.s.=20 div)
±0.1% f.s.
2 μs to 2 s
Measurement
resolution
Count mode: 2000LSB/div (f.s. = 20 div)
Except for count mode and power frequency mode: 500LSB/div (f.s. = 20 div)
Power frequency mode: 100LSB/div (f.s. = 20 div)
Response time
Within 40 μs + sampling interval of instrument in which the module is installed.
Input voltage range
±10 V, ±20 V, ±50 V, ±100 V, ±200 V, ±400 V
Threshold value
±10 V range: −10 V to +10 V variable (0.1-V step)
±20 V range: −20 V to +20 V variable (0.2-V step)
±50 V range: −50 V to +50 V variable (0.5-V step)
±100 V range: −100 V to +100 V variable (1-V step)
±200 V range: −200 V to +200 V variable (2-V step)
±400 V range: −400 V to +400 V variable (5-V step)
Slope
Rising, falling (Frequency, rotation speed, power frequency, count mode)
Level
High, Low (Duty ratio mode, pulse width mode)
Hold
Frequency mode/Rotation speed mode: ON, OFF (1 Hz, 0.5 Hz, 0.2 Hz, 0.1 Hz)
When OFF is selected: If the next measurement value is not determined within the waiting
time period, the frequency or the rotation speed will be determined
using the value that is calculated based on the time period
between the timing of the last measurement and the timing of the
sampling. If these calculated values are less than the specified
values, then use 0.
Smoothing
OFF, ON (The permissible smoothing frequency is up to 10 kHz.)
(Frequency mode, rotation speed mode)
Low-pass filter
OFF, 5, 50, 500, 5 k, 50 kHz
Input coupling
DC, AC (Low frequency cut-off in AC-coupled mode: 7 Hz)
Frequency dividing
function
Setting range: 1 to 4096, by one step (Frequency, rotation speed, count mode)
Counting start timing
Start, trigger (Count mode)
Count over handling
Hold, back (Count mode)
Operating
temperature and
humidity
As per Memory HiCorder in which Model 8970 is installed
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......